Figure 25 - Continuous Transient Operation Current Waveform
Pulse
In this mode, the electronic load generates a transient pulse of programmable width when pulse
transient operation is in effect.
In pulse mode, you can set A/B level, A/B width, and A/B slew rate via the mainframe keypad. The
electronic load will automatically switch to A level after maintaining A width time. Then it will
switch to B level. The electronic load will not switch to A level again until the instrument receives
the pulse signal. The following figure shows the current waveform in pulse transient operation.
Figure 26 - Pulse Transient Operation